23
DL-00096 Rev. 7
Immunity test
Electrostatic
discharge(ESD)
IEC61000-4-2
Electricfast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
IEC 60601 test
level
±6kVcontact
±8kVair
±2kVfor
power supply
lines
±1kVfor
input/output
lines
±1kV
differential mode
±2kVcommon
mode
Electromagnetic
environment/guidance
Floorsshouldbewood,
concrete,orceramictile.
Iffloorsarecoveredwith
synthetic material, the
relativehumidityshould
beatleast30%.
Mainspowerquality
should be that of a typical
commercial or hospital
environment.
Mainspowerquality
should be that of a typical
commercial or hospital
environment.
Guidance and manufacturer’s declaration: electromagnetic immunity
ThePortableOxygenConcentratorisintendedforuseintheelectromagnetic
environmentspecifiedbelow.Thecustomerortheuseroftheconcentratorshould
assurethatitisusedinsuchanenvironment.
Compliance
level
±15kVcontact
±8kVair
±2kVfor
power supply
lines
±1kVfor
input/output
lines
±1kV
differential mode
±2kVcommon
mode
DL-00096 rev 7.indd 23 6/1/2009 4:48:35 PM