Delphi DL-00183 Manual de usuario Pagina 29

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 40
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 28
17
DL-00096 Rev. 7
Alarm Message
CheckVents
LowBattery
XX:Service!*
*Value:21-50
Description
Deviceisunabletomaintain
oxygenpurity.
Estimatedbatterylifeless
than17minutes.
Servicerequired.
Action
Besureairinlet/outlet
hasnotbeenblocked.If
alarm persists, contact
CustomerService.
Chargethebatterypack
by plugging in to power
supply.
Note: The message will be
automatically cleared when
plugged in to power supply.
ContactCustomer
Service.
High Priority Alarms
Note: These alarms will allow the Portable Oxygen Concentrator to continue operating.
Alarm Message
CheckCannula
XX:Service*
*Value:51-70
Description
Nobreathdetectedfor5
minutes.
Servicerequired.
Action
Checkthecannulacon-
nection.Besureto
breathethroughnose.If
alarm persists, contact
CustomerService.
Note: The message will be
automatically cleared when
breathing is detected.
ContactCustomer
Service.
Medium Priority Alarms
DL-00096 rev 7.indd 17 6/1/2009 4:48:34 PM
Vista de pagina 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 ... 39 40

Comentarios a estos manuales

Sin comentarios